All guides  ›  Internal quality control

Why run controls at more than one concentration?

A control level tells you about the behaviour of the method at that concentration. Running more than one level can reveal a change at one part of the measuring range while results at another level still look stable.

Worked example at a glance

20 units

Low control target

100 units

High control target

Run 6

First result above the low control’s limit

Choose levels around the results you need to monitor

Start with the concentrations at which patients’ results are used and where a change in analytical performance could matter. A control near a relevant decision concentration can help monitor that part of the range. Another level may help detect changes elsewhere. The choice also depends on the method, control materials available and the laboratory’s QC plan. [1] [2]

Using a second concentration answers a different question from measuring the same control twice. A repeat tells you more about that material at its own concentration. Another level adds a different view of the measuring range. Give each level its own target and limits, based on the procedure and conditions it is meant to monitor.

Read the two charts together

A fictional laboratory monitors one method with controls near 20 and 100 units. It records an analyser service between runs 5 and 6. From run 6, the low control results are 22.0, 22.2 and 22.1 units, each above its established upper limit of 21.8. The high control continues to give results close to 100 units.

After a service between runs 5 and 6, three low-control results are above their three-SD limit, while high-control results stay close to their target. Each chart is expressed in standard deviations from its own target.
Both charts use SDs from their own targets. The shaded area marks the runs after service. The low control crosses its upper limit while the high control remains near its target. Open the figure full size.

The chart displays each result as a number of SDs from its own target. This makes the patterns easier to compare even though the concentrations and SDs differ. The high control gives a useful observation about its own level. It does not resolve the signal at the low level.

A useful control level needs to be measurable reliably and suitable for the procedure. Its behaviour may differ from a patient specimen, so the level and material both matter. If the available low control sits well away from the concentration used for a decision, record the limitation and consider what other evidence covers that part of the range. [2]

Decide which part of the range needs attention

The pattern gives the laboratory a reason to look at the low level and the events around the service. Possible explanations include the control material, its preparation, calibration or method behaviour at that concentration. The chart alone cannot distinguish these. Comparing the records at both levels helps define the question to investigate. [1] [2]

Control levelEstablished target and limits
LowTarget 20 units · limits 18.2–21.8
HighTarget 100 units · limits 94–106
Fictional laboratory values. The two levels have different targets and SDs.

Keep the measured units and the control level in the review. A similar number of SDs on two charts can represent different absolute differences. Whether a change could matter for patient results depends on the relevant concentrations, the measurement procedure and the evidence from the follow-up.

The charts are aligned by run number. That lets you see that the low-level change and the stable high-level observations occur in the same runs. Comparing two charts from different periods would weaken that interpretation.

Describe what each level showed

The two levels provide evidence at two points. They do not sample every concentration in between. Where performance near another decision level matters, consider whether the available controls and other evidence give you an adequate view of that part of the range. Record why you selected the levels and what they help you monitor. [1] [2]

Example wordingAfter service, the low control exceeded its upper limit in runs 6–8. The high control remained near its own target. The signal was confined to the low control in the data reviewed, and its cause and relevance to patient results require follow-up.

Member Learning

Practise reading a control chart

Understanding Internal Quality Control (IQC) takes you through the chart, the changes that need attention and the laboratory’s response.

Sources and related reading

Read why a chart can signal within limits · Read about fitness for purpose in analytical performance specifications

By Stephen MacDonald · Suggest a correction · Back to Find a guide